Slot-Dependent Wireless Power Transfer System for MBAN Applications

2020

Closed access Metadata (CSL-JSON)
DOI 10.1109/icce-berlin50680.2020.9352185
Data confidence 85 high
How this score is calculated

It is built from checkable properties of the record itself: whether the article has a DOI, how many independent sources confirmed its data, whether the required fields are filled in, and how recent the latest confirmation is. Disagreement between sources and a missing identifier lower it. A value of 80 or above means the record was published without further checking; 60 to 79 means it is published at reduced confidence, with a manual check advisable. A record verified by an editor by hand is never pushed below 90 by a recalculation. This scores our data about the article, not the article itself.

Abstract

This paper presents for the first time the wireless power transfer (WPT) system that is fully dependent on the etched slot on the ground plane. At first, the defected ground structure-based resonator with the circuit area of 20-by-20 sq. mm and a rectangular defect was proposed. Then, it was changed to three rectangular slots to achieve resonance at the desired frequency. Subsequently, the resonators with optimized parameters were aptly utilized to design the WPT system that works at 2.4 GHz. The operating frequency complies with the practical medical body area network band and, therefore, can be employed to the biomedical implants. It has been demonstrated that the perfect coupling can be observed at a distance of 10 mm. As a consequence, at this distance the obtained simulation results of WPT show 67% efficiency and the figure of merit of 0.335.