Wireless Power and Information Transfer System Employing DGS based Resonators

2026

Closed access Metadata (CSL-JSON)
DOI 10.23919/nrsm68586.2026.11550883
Data confidence 70 medium
How this score is calculated

It is built from checkable properties of the record itself: whether the article has a DOI, how many independent sources confirmed its data, whether the required fields are filled in, and how recent the latest confirmation is. Disagreement between sources and a missing identifier lower it. A value of 80 or above means the record was published without further checking; 60 to 79 means it is published at reduced confidence, with a manual check advisable. A record verified by an editor by hand is never pushed below 90 by a recalculation. This scores our data about the article, not the article itself.

Abstract

we present design and evaluation of a wireless power transfer (WPT) system with a small size of $\mathbf{2 5} \boldsymbol{\times} \mathbf{2 5} \mathrm{mm}^{\mathbf{2}}$ aimed at application in implantable medical devices. The initial step in the design of the WPT entails the development of a compact single-band defected ground structure, in this case patterned square defects, based resonator. Then two identical resonators are coupled at an optimal separation distance of 26mm to form the WPT system. The prototyped WPT achieves an efficiency of $\mathbf{6 6 . 4 \%}$ at the design frequency of $\mathbf{4 3 3 ~ M H z}$. The WPT’s data transmission capabilities are evaluated based on bit-error-rate (BER) using universal software radio peripheral configured through LabVIEW software. Specifically, BER is measured at varying separation distances between the resonators, reaching value of 0.026 at 26mm.