EXPERIMENTAL STUDIES OF THE DISTRIBUTIONS OF THE DEPTH INTENSITY OF THE ELECTRON BEAM IN THE BODY OF THE TREATED PRODUCTS

2018

Full text Open access: Hybrid Metadata (CSL-JSON)
DOI 10.52676/1729-7885-2018-4-97-99
Data confidence 70 medium
How this score is calculated

It is built from checkable properties of the record itself: whether the article has a DOI, how many independent sources confirmed its data, whether the required fields are filled in, and how recent the latest confirmation is. Disagreement between sources and a missing identifier lower it. A value of 80 or above means the record was published without further checking; 60 to 79 means it is published at reduced confidence, with a manual check advisable. A record verified by an editor by hand is never pushed below 90 by a recalculation. This scores our data about the article, not the article itself.

Abstract

The measurements of geometrical and physical parameters of the processed medical products were carried out. The density of samples was measured both in industrial packaging and in individual products. Experimental studies on the passage of electrons with an energy of 2 MeV through medical products: cotton wool, syringes. The absorption coefficients of the electron beam for two materials (cotton wool, syringes) are obtained. It was found that when irradiating products from the sides there is a greater uniformity of dose distribution in samples of cotton wool and syringes. From the results of the experiment it follows that such irradiation leads to the fact that the unevenness of the dose of the impact on the thickness of the sample for cotton wool is about 5 %, and for syringes – 3 %.