The measurements of geometrical and physical parameters of the processed medical products were carried out. The density of samples was measured both in industrial packaging and in individual products. Experimental studies on the passage of electrons with an energy of 2 MeV through medical products: cotton wool, syringes. The absorption coefficients of the electron beam for two materials (cotton wool, syringes) are obtained. It was found that when irradiating products from the sides there is a greater uniformity of dose distribution in samples of cotton wool and syringes. From the results of the experiment it follows that such irradiation leads to the fact that the unevenness of the dose of the impact on the thickness of the sample for cotton wool is about 5 %, and for syringes – 3 %.