OPTICAL DIAGNOSTICS SYSTEM AT THE INURA ACCELERATOR

2023

Full text Open access: Hybrid Metadata (CSL-JSON)
DOI 10.52676/1729-7885-2023-1-77-82
Data confidence 70 medium
How this score is calculated

It is built from checkable properties of the record itself: whether the article has a DOI, how many independent sources confirmed its data, whether the required fields are filled in, and how recent the latest confirmation is. Disagreement between sources and a missing identifier lower it. A value of 80 or above means the record was published without further checking; 60 to 79 means it is published at reduced confidence, with a manual check advisable. A record verified by an editor by hand is never pushed below 90 by a recalculation. This scores our data about the article, not the article itself.

Abstract

A system has been created for extracting and transporting light radiation from the irradiation chamber and the protective box of the INURA accelerator. Light outside the protective box is recorded using compact spectrometers, photomultipliers, photodetectors in the range from 300 nm to 2.1 µm. At present, the accelerator operates in the ion acceleration mode; the number of problems to be solved should increase when switching to the electron acceleration mode. The system was tested in the registration of light radiation from the interelectrode gap of the diode assembly. Lines of a hydrogen atom and an oxygen atom associated with the decomposition of absorbed water vapor and the discharge over the dielectric surface at the anode of the high-current accelerator diode have been registered.