It is built from checkable properties of the record itself: whether the article has a DOI, how many independent sources confirmed its data, whether the required fields are filled in, and how recent the latest confirmation is. Disagreement between sources and a missing identifier lower it. A value of 80 or above means the record was published without further checking; 60 to 79 means it is published at reduced confidence, with a manual check advisable. A record verified by an editor by hand is never pushed below 90 by a recalculation. This scores our data about the article, not the article itself.

Abstract

The results of research on the application of the method of measurement of gross alpha-, beta-activity in water using the liquid scintillation counter (LSC) Quantulus 1220 are presented. Basic procedures of preparation and spectrometric measurement of aqueous samples were studied and performed. The detection efficiency of alpha-, beta-particles depending on quenching and the optimum alpha-, beta-radiation separation parameter for the method used were determined. Testing and validation of the technique was carried out on model solutions with different concentrations and composition of alpha- and beta-emitters, as well as by participating in interlaboratory comparisons. The results demonstrated that the use of LSS allows for a sufficiently fast and accurate estimation of the gross alpha- and beta-activity in water below the regulated limits without the use of labor-intensive sample preparation.