Analysis of flow resistance in bundles of power plant condensers

2019

Full text Open access: Gold Metadata (CSL-JSON)
DOI 10.1051/e3sconf/201910000071
Data confidence 70 medium
How this score is calculated

It is built from checkable properties of the record itself: whether the article has a DOI, how many independent sources confirmed its data, whether the required fields are filled in, and how recent the latest confirmation is. Disagreement between sources and a missing identifier lower it. A value of 80 or above means the record was published without further checking; 60 to 79 means it is published at reduced confidence, with a manual check advisable. A record verified by an editor by hand is never pushed below 90 by a recalculation. This scores our data about the article, not the article itself.

Abstract

Shell-side pressure drop is a very important variable in the successful design of the condensers. The prediction of this pressure drop through the horizontal tube banks, with condensation, has long been a problem facing design engineers. Low pressure drop is a requirement in designing condensers for power plants. The paper presents a comparison of the various correlation to determine the pressure drop in the tube bundle. It is an important element for the verification of numerical simulations. Analysis of flow resistance for power plant condenser were made.