DEVICE FOR MEASURING THE THERMAL CONDUCTIVITY OF ORGANIC SUBSTANCES AT LOW TEMPERATURES

2018

Closed access Metadata (CSL-JSON)
DOI 10.1615/jenhheattransf.2018026383
Data confidence 70 medium
How this score is calculated

It is built from checkable properties of the record itself: whether the article has a DOI, how many independent sources confirmed its data, whether the required fields are filled in, and how recent the latest confirmation is. Disagreement between sources and a missing identifier lower it. A value of 80 or above means the record was published without further checking; 60 to 79 means it is published at reduced confidence, with a manual check advisable. A record verified by an editor by hand is never pushed below 90 by a recalculation. This scores our data about the article, not the article itself.

Abstract

An installation designed for measuring the thermal conductivity of organic materials, including biological tissues, is described at temperatures below room temperature up to 100 K using the method of stationary heat flow. The advantages of the installation include the ability to set the temperature regime of freezing the sample, smoothly changing the rate of temperature front movement in the sample during cooling. The output of the device to the preset stationary mode and the measurements are carried out automatically. The values of the coefficient of thermal conductivity of the muscle tissue calculated on the basis of the measurement results when the samples are cooled in the temperature range from 280 K to 100 K are in good agreement with the literature data.