Error of a Temperature Probe for Cancer Ablation Monitoring Caused by Respiratory Movements: Ex Vivo and In Vivo Analysis

2016

Full text Open access: Green Metadata (CSL-JSON)
DOI 10.1109/jsen.2016.2574959
Data confidence 85 high
How this score is calculated

It is built from checkable properties of the record itself: whether the article has a DOI, how many independent sources confirmed its data, whether the required fields are filled in, and how recent the latest confirmation is. Disagreement between sources and a missing identifier lower it. A value of 80 or above means the record was published without further checking; 60 to 79 means it is published at reduced confidence, with a manual check advisable. A record verified by an editor by hand is never pushed below 90 by a recalculation. This scores our data about the article, not the article itself.

Abstract

Hyperthermal techniques are spreading as an alternative to conventional surgery for cancer removal. A real-time temperature feedback can be used to adjust the treatment settings, in order to improve the clinical outcomes. In this paper, we experimentally assessed the feasibility for distributed temperature monitoring of a custom probe, which consists of a needle embedding six fiber Bragg gratings (FBGs). Since FBGs are also sensitive to strain, we focused on the analysis of the measurement error (artifact) caused by respiratory movements. We assessed the artifact both onex vivopig liver and lung (by mimicking the movement of these organs caused by respiration) and onin vivotrial on pig liver. Lastly, we proposed an algorithm to detect and minimize the artifact duringex vivoliver laser ablation. During bothex vivoandin vivotrials, the probe insertion within the organ was easy and safe. The artifact was significant (up to 3 °C), but the correction algorithm allows minimizing the error. The main advantages of the proposed probe are: 1) spatially resolved temperature monitoring (in six points of the tissue by inserting a single needle) and 2) the needle is magnetic resonance (MR)-compatible, hence can be used during MR-guided procedure. Even if the model is close to humans, further trials are required to investigate the feasibility of the probe for clinical applications.