SPME GC/MS determination of organochlorine pesticides in water samples

2013

Full text Open access: Gold Metadata (CSL-JSON)
DOI 10.15328/chemb_2013_19-16
Data confidence 70 medium
How this score is calculated

It is built from checkable properties of the record itself: whether the article has a DOI, how many independent sources confirmed its data, whether the required fields are filled in, and how recent the latest confirmation is. Disagreement between sources and a missing identifier lower it. A value of 80 or above means the record was published without further checking; 60 to 79 means it is published at reduced confidence, with a manual check advisable. A record verified by an editor by hand is never pushed below 90 by a recalculation. This scores our data about the article, not the article itself.

Abstract

Headspace solid phase microextraction (HS-SPME) in combination with gas chromatography and mass-spectrometry (GC/MS) was studied for analysis of water samples. The organochlorine pesticides (OCPs), p,p'-DDT, p,p'-DDD, and p,p'-DDE were collected and analyzed by GC/MS. To select of effective fiber coatings four types of SPME fibers were examined and compared. The parameters effecting the efficiency of HS-SPME such as extraction and pre-incubation time and extraction temperature, effect of solvent nature, ionic strength were studied to obtain optimal parameters. The method was developed using spiked water samples in a concentration range 10 - 500 ng/L. The calibration curve was linear over the studied concentration range with r≥0.9925. The detection limits varied from 1.57 to 2.08 ng/L. An authentic water samples from contaminated lake with OCPs were analyzed by developed method.