Temperature Monitoring during Radio Frequency Thermal Ablation Treatment on Ex Vivo perfused organ by Fiber Bragg Grating Sensors

2018

Closed access Metadata (CSL-JSON)
DOI 10.1109/memea.2018.8438730
Data confidence 70 medium
How this score is calculated

It is built from checkable properties of the record itself: whether the article has a DOI, how many independent sources confirmed its data, whether the required fields are filled in, and how recent the latest confirmation is. Disagreement between sources and a missing identifier lower it. A value of 80 or above means the record was published without further checking; 60 to 79 means it is published at reduced confidence, with a manual check advisable. A record verified by an editor by hand is never pushed below 90 by a recalculation. This scores our data about the article, not the article itself.

Abstract

In this work, we report on the integration of fiber Bragg gratings sensors in a commercial radio frequency probe for real-time temperature monitoring during thermal ablation of biological tissues. The experiments were conducted on ex-vivo pig liver and bovine kidney tissues and the results confirm that the proposed setup is able to conduct a multi-point real-time temperature monitoring during the RF discharges with a temperature resolution of 0.1 °C and a minimum spatial resolution of 1 mm. During these experiments the blood perfusion was implemented by injecting water at a temperature of 37 °C inside the organ under examination. This solution aimed at simulating the Radio Frequency Thermal Ablation (RFTA) treatments made in clinical practice.