Temperature monitoring during thermal ablation on ex-vivo organs by Fiber Bragg gratings

2016

Closed access Metadata (CSL-JSON)
DOI 10.1109/icsens.2016.7808544
Data confidence 70 medium
How this score is calculated

It is built from checkable properties of the record itself: whether the article has a DOI, how many independent sources confirmed its data, whether the required fields are filled in, and how recent the latest confirmation is. Disagreement between sources and a missing identifier lower it. A value of 80 or above means the record was published without further checking; 60 to 79 means it is published at reduced confidence, with a manual check advisable. A record verified by an editor by hand is never pushed below 90 by a recalculation. This scores our data about the article, not the article itself.

Abstract

In this paper we report on the application of Fiber Bragg gratings (FBGs) for temperature monitoring during radiofrequency ablation (RFA) for tumor treatment. A proper configuration with several FBG sensors deployed close to the region reached by RF discharges has been developed to monitor the necrotized area. Our ex-vivo experiments on animal kidney and liver confirm that we were able to monitor the temperature with a resolution of 0.1 °C during a series of different and consecutive RF discharges with the laparoscopic bipolar radiofrequency device Habib 4x.