Thermal gradient estimation with fiber-optic chirped FBG sensors: Experiments in biomedical applications

2017

Closed access Metadata (CSL-JSON)
DOI 10.1109/icsens.2017.8234119
Data confidence 70 medium
How this score is calculated

It is built from checkable properties of the record itself: whether the article has a DOI, how many independent sources confirmed its data, whether the required fields are filled in, and how recent the latest confirmation is. Disagreement between sources and a missing identifier lower it. A value of 80 or above means the record was published without further checking; 60 to 79 means it is published at reduced confidence, with a manual check advisable. A record verified by an editor by hand is never pushed below 90 by a recalculation. This scores our data about the article, not the article itself.

Abstract

The measurement of short-scale (sub-millimeter) thermal gradients within 1-5 cm range can have a significant impact on several biomedical applications, including the measurement of effectiveness of thermal ablation, the estimation of blood pressure thermal differences during angioplasty, and fractional flow reserve. In this work, we propose and experimentally validate a technique to measure a thermal gradient with a fiber-optic sensor based on a linearly chirped fiber Bragg grating (LCFBG). The LCFBG spectrum is measured with a white-light setup, and a demodulation of the grating using the coupled mode theory model is used to estimate the thermal gradient. Experiments are reported in a calibration setup for a linear gradient, achieving <;1° C error, and in a laser ablation with a Gaussian-shaped thermal gradient.