Development and optimization of the production technology of large-size position-sensitive detectors

2014

Closed access Metadata (CSL-JSON)
DOI 10.1134/s0020441214040083
Data confidence 70 medium
How this score is calculated

It is built from checkable properties of the record itself: whether the article has a DOI, how many independent sources confirmed its data, whether the required fields are filled in, and how recent the latest confirmation is. Disagreement between sources and a missing identifier lower it. A value of 80 or above means the record was published without further checking; 60 to 79 means it is published at reduced confidence, with a manual check advisable. A record verified by an editor by hand is never pushed below 90 by a recalculation. This scores our data about the article, not the article itself.

Abstract

Results of development and optimization of the production technology of large-size semiconductor position-sensitive X-ray detectors, based on Si(Li)- p - i - n structures and intended for tomographic and environmental problems, are described. The detectors with a 50 × 50 × 2-mm-size sensitive area with 8, 16, and 32 strips were produced. With a voltage of 100–500 V, the energy resolution is 50 and 75–80 keV for electrons with an energy of 1 MeV and for α particles with an energy of 7.65 MeV, respectively, when the dark current is 0.5–1 μA, the capacitance value is 300 pF, and noise is 40 keV.