Calibration of industrial platinum resistance thermometers in thermostats (-40 °C to 420 °C)

2019

Full text Open access: Gold Metadata (CSL-JSON)
DOI 10.1088/0026-1394/56/1a/03003
Data confidence 70 medium
How this score is calculated

It is built from checkable properties of the record itself: whether the article has a DOI, how many independent sources confirmed its data, whether the required fields are filled in, and how recent the latest confirmation is. Disagreement between sources and a missing identifier lower it. A value of 80 or above means the record was published without further checking; 60 to 79 means it is published at reduced confidence, with a manual check advisable. A record verified by an editor by hand is never pushed below 90 by a recalculation. This scores our data about the article, not the article itself.

Abstract

The COOMET.T-S2 comparison is a regional comparison in which the national metrological institutes of Moldova, Georgia, Kyrgyzstan, Kazakhstan, Azerbaijan, Bosnia and Herzegovina and Turkey participate. Two standard platinum resistance thermometers (100 ohm), manufactured at VNIIM, was used as a transfer standards. National Metrology Institute (NMI) of Republic of Moldova is a pilot laboratory in this comparison. The purpose of the comparisons is to verify equivalence between the participating laboratories in the calibration of industrial platinum resistance thermometers by direct comparison in the range from minus 40 °C to 420 °C. The comparison results are used to confirm the calibration and measurement capabilities in this range. Main text To reach the main text of this paper, click on Final Report . Note that this text is that which appears in Appendix B of the BIPM key comparison database kcdb.bipm.org/ . The final report has been peer-reviewed and approved for publication by the CCT, according to the provisions of the CIPM Mutual Recognition Arrangement (CIPM MRA).