Measurements of temperature during thermal ablation treatments on ex vivo liver tissue using fiber Bragg grating sensors

2017

Closed access Metadata (CSL-JSON)
DOI 10.1109/i2mtc.2017.7969742
Data confidence 70 medium
How this score is calculated

It is built from checkable properties of the record itself: whether the article has a DOI, how many independent sources confirmed its data, whether the required fields are filled in, and how recent the latest confirmation is. Disagreement between sources and a missing identifier lower it. A value of 80 or above means the record was published without further checking; 60 to 79 means it is published at reduced confidence, with a manual check advisable. A record verified by an editor by hand is never pushed below 90 by a recalculation. This scores our data about the article, not the article itself.

Abstract

In this work, we propose the use of fiber Bragg grating (FBG) sensor arrays for real-time temperature monitoring during RadioFrequency Thermal Ablation (RFTA) tumor treatment. Real-time temperature monitoring of RF-treatment to the tumors lesion of an organ could prove to be highly beneficial for intra-operative surgical planning and subsequently ensure a successful outcome of a thermo-ablation procedure. An adequate configuration was designed in order to create a thermal multipoint map. In particular, the RF probe of a commercial medical device was properly packaged with FBG sensors. In order to discern the treatment areas as accurately as possible, a second 3.5 cm long array, consisting of several FBGs was also employed. The experiments were conducted on ex vivo animal liver tissues and results confirm that we were successfully able to conduct a multipoint measurement and to distinguish between different and consecutive RF discharges with a temperature resolution of 0.1 °C and a minimum spatial resolution of 5mm.