Quantification of atomic force microscopy tip and sample thermal contact

2019

Full text Open access: Green Metadata (CSL-JSON)
DOI 10.1063/1.5097862
Data confidence 70 medium
How this score is calculated

It is built from checkable properties of the record itself: whether the article has a DOI, how many independent sources confirmed its data, whether the required fields are filled in, and how recent the latest confirmation is. Disagreement between sources and a missing identifier lower it. A value of 80 or above means the record was published without further checking; 60 to 79 means it is published at reduced confidence, with a manual check advisable. A record verified by an editor by hand is never pushed below 90 by a recalculation. This scores our data about the article, not the article itself.

Abstract

A thermal conduction measurement device was fabricated, consisting of a silicon dioxide membrane with integrated thermal sensors (Pt resistance heater/thermometer and Pt–Au thermocouples) using MEMS technology. Heat transfer between the heated device and a number of unused atomic force microscope and scanning thermal microscope probes was measured. Changes in thermal conduction related to changes in the tip shape resulting from initial contact were observed. The sensors were fabricated by electron beam lithography and lift-off followed by local subtractive processing of a Pt–Au multilayer to form Pt heater–resistance thermometer elements and Pt–Au thermocouples. Thermal isolation from the silicon substrate was provided by dry release of the supporting 50 nm thick SiO2 membrane using an isotropic SF6 inductively coupled plasma etch. The high thermal isolation of the sample combined with the sensitivity of the temperature sensors used allowed the detection of thermal conduction between the tip and the sample with high precision. The measured temperature range of the Pt resistor was 293–643 K. The measured thermal resistance of the membrane was 3 × 105 K/W in air and 1.44 × 106 K/W in vacuum. The tip contact resistance was measured with a noise level of 0.3g0 T at room temperature, where g0 is the thermal resistance quantum.