GaAs Schottky Barrier Detectors for Alpha-Particle Spectrometry at Temperatures up to 120°C

2018

Closed access Metadata (CSL-JSON)
DOI 10.1134/s106378501810019x
Data confidence 70 medium
How this score is calculated

It is built from checkable properties of the record itself: whether the article has a DOI, how many independent sources confirmed its data, whether the required fields are filled in, and how recent the latest confirmation is. Disagreement between sources and a missing identifier lower it. A value of 80 or above means the record was published without further checking; 60 to 79 means it is published at reduced confidence, with a manual check advisable. A record verified by an editor by hand is never pushed below 90 by a recalculation. This scores our data about the article, not the article itself.

Abstract

GaAs Schottky barrier detectors for α particle spectrometry have been tested. Detectors had an input window area of 80 mm 2 and a working barrier layer thickness of 40–50 μm. The energy resolution (FWHM) measured on 5.499 MeV α line of 238 Pu source amounted to 17.5 keV at a generator peak width of 7.8 keV. In the measurements on a 226 Ra source, detectors showed linear response and nearly 100% charge collection efficiency at reverse bias above 65 V for all energies of α particles emitted from the source. Tests for thermal stability showed that the proposed detectors can be used in alpha-spectrometry of radionuclides at temperatures up to 120°C.