Case report: Transcatheter tricuspid valve-in-valve implantation using novel balloon-expandable aortic valve with 1 year follow-up

2023

Full text Open access: Gold Metadata (CSL-JSON)
DOI 10.3389/fcvm.2023.1152280
Data confidence 70 medium
How this score is calculated

It is built from checkable properties of the record itself: whether the article has a DOI, how many independent sources confirmed its data, whether the required fields are filled in, and how recent the latest confirmation is. Disagreement between sources and a missing identifier lower it. A value of 80 or above means the record was published without further checking; 60 to 79 means it is published at reduced confidence, with a manual check advisable. A record verified by an editor by hand is never pushed below 90 by a recalculation. This scores our data about the article, not the article itself.

Abstract

Generally, the dysfunction or failure of bioprosthetic heart valves (BHVs) is managed by replacement surgery. In the case of tricuspid valve dysfunction, re-do surgery is rarely attempted because of the critically high risk of developing pulmonary hypertension, pulmonary embolism, and intraoperative mortality. Hence, transcatheter tricuspid repair and replacement procedures are preferred. More recently, transcatheter valve-in-valve (ViV) treatments have gained importance because of their less invasiveness, especially for patients with prior surgeries. Encouraging evidence of the safety and effectiveness of a novel balloon-expandable (BE) transcatheter heart valve (THV)-the Myval THV-has been reported for ViV procedures. Here, we present a case-series of 5 patients, in whom tricuspid ViV procedure was performed using BE Myval THV, implanted supra-annularly by anchoring onto the deteriorated BHV. This case-series details the procedural steps to prevent in-hospital adverse events and early (30-day) mortality and the challenges during tricuspid ViV interventions.