Fiber Bragg Grating for Temperature Monitoring During Medical Radiofrequency Treatments

2016

Full text Open access: Gold Metadata (CSL-JSON)
DOI 10.1016/j.proeng.2016.11.354
Data confidence 70 medium
How this score is calculated

It is built from checkable properties of the record itself: whether the article has a DOI, how many independent sources confirmed its data, whether the required fields are filled in, and how recent the latest confirmation is. Disagreement between sources and a missing identifier lower it. A value of 80 or above means the record was published without further checking; 60 to 79 means it is published at reduced confidence, with a manual check advisable. A record verified by an editor by hand is never pushed below 90 by a recalculation. This scores our data about the article, not the article itself.

Abstract

In this work, it is proposed the integration of fiber optic sensors with commercial medical instrumentation, for temperature monitoring during tumor radiofrequency thermal ablation (RFTA) treatments. The heat propagation over the area under treatment is measured by means of a specific configuration, consisting of several Fiber Bragg Grating (FBG) sensors properly bounded to the RFTA probe and being positioned close to the region reached by RF electrodes. Our ex-vivo experiments on ex vivo animal liver and kidney, confirm that we were able to monitor in real time the temperature increase during different and consecutive RF discharges performed with the laparoscopic bipolar device Habib 4x, with a temperature resolution of 0.1̊C and a spatial resolution of 5mm.