Selection of a calibration system for relative gravimeters and testing of the processing using the example of the Zhetygen calibration baseline in Kazakhstan

2024

Full text Open access: Hybrid Metadata (CSL-JSON)
DOI 10.1007/s40328-024-00454-x
Data confidence 70 medium
How this score is calculated

It is built from checkable properties of the record itself: whether the article has a DOI, how many independent sources confirmed its data, whether the required fields are filled in, and how recent the latest confirmation is. Disagreement between sources and a missing identifier lower it. A value of 80 or above means the record was published without further checking; 60 to 79 means it is published at reduced confidence, with a manual check advisable. A record verified by an editor by hand is never pushed below 90 by a recalculation. This scores our data about the article, not the article itself.

Abstract

Abstract This article consolidates and organizes modern methods for calibrating relative gravimeters, aiming to streamline the selection and implementation of effective calibration systems, particularly in Kazakhstan. The concept of the calibration function and various methods for its determination are presented. Practical implementations of laboratory methods relevant to modern high-precision relative gravimeters are discussed, including the tilt, moving mass, artificial acceleration, and line calibration methods. Data processing through least squares adjustment at the calibration line and an overview of existing software packages for gravity observation equalization are explored. The article also covers existing horizontal and vertical calibration systems in different countries, detailing their main characteristics and schematics. Finally, an estimation of scale factors for five years of measurements with Scintrex CG-5 gravimeters at the Zhetygen calibration line in Kazakhstan is provided.