Development and Finite Element Analysis of a Titanium Bone Plate with a Localized Porous Structure for Osteosynthesis of the Radial Shaft

Applied Sciences, 2026

Full text Open access: Gold Metadata (CSL-JSON)
DOI 10.3390/app16178405
Data confidence 85 high
How this score is calculated

It is built from checkable properties of the record itself: whether the article has a DOI, how many independent sources confirmed its data, whether the required fields are filled in, and how recent the latest confirmation is. Disagreement between sources and a missing identifier lower it. A value of 80 or above means the record was published without further checking; 60 to 79 means it is published at reduced confidence, with a manual check advisable. A record verified by an editor by hand is never pushed below 90 by a recalculation. This scores our data about the article, not the article itself.

Abstract

Traditional bone plates provide reliable fracture fixation; however, their high stiffness can cause stress shielding, thereby reducing mechanical stimulation of the bone tissue and slowing its regeneration. The biomechanical performance of a titanium plate with a localized porous structure for osteosynthesis of the radial shaft was investigated in this study. Three designs were considered: a solid plate and two plates with localized porous regions measuring 10 × 10 mm and 10 × 15 mm. The finite element analysis of the bone–plate system was performed using MSC Patran/Nastran, with rigid fixation of the proximal end of the bone and sequential application of an axial compressive load of 100 N, bending, and torsion with a moment of 1 N·m. Biomechanical performance was evaluated based on von Mises equivalent stress, fragment displacement (FD), interfragmentary movement (IFM), interfragmentary strain (IFS), and strain energy density (SED). To confirm the manufacturability of the design, the plate was fabricated from Ti–6Al–4V alloy using laser powder bed fusion (LPBF), and the geometry of the porous structure was verified by scanning electron microscopy. The results showed that the localized porous structure altered the load distribution between the plate and the bone, resulting in an increase in local stresses in the bone under the investigated loading conditions. These changes indicate an alteration in the mechanical environment within the bone, which may potentially affect conditions related to fracture healing. Among the investigated configurations, the plate with a 10 × 10 mm porous insert demonstrated the most balanced mechanical characteristics in terms of stresses in the implant, stress distribution in the bone, and structural stability.