It is built from checkable properties of the record itself: whether the article has a DOI, how many independent sources confirmed its data, whether the required fields are filled in, and how recent the latest confirmation is. Disagreement between sources and a missing identifier lower it. A value of 80 or above means the record was published without further checking; 60 to 79 means it is published at reduced confidence, with a manual check advisable. A record verified by an editor by hand is never pushed below 90 by a recalculation. This scores our data about the article, not the article itself.

Abstract

Abstract This article presents studies on the dew point/frost temperature generator using the single-pressure (1-P) and single-temperature (1-T) method first developed by KazStandard to improve the calibration accuracy of dew point hygrometers. The method of applying the dew point temperature scale is based on the supply of humid air with high instability through a thermostatic saturator, which enables increased measurement accuracy. The characteristics include a study of the saturator's efficiency, the stability of the dew point temperature reproduction and a comparison with a calibrated dew point hygrometer. The study was conducted in the dew point temperature range from −50°C to 20°C, and the sources of measurement uncertainty were analyzed.