The dose distribution around the applicator in HDR brachytherapy on GammaMed+ unit using a 192Ir source was evaluated using GafChromic films.The film was calibrated using a TrueBeam linear accelerator.The experimental dose distribution results match the planned Eclipse system and also indicate that the discrepancy between the planned system and measured isodose values is approximately ±1.2%.