Аңдатпа
Abstract The work deals with the selection of conditions for the preparation of biomonitor mosses using a microwave decomposition system for the determination of heavy metals entering them during transboundary transfer from atmospheric air. The completeness of mineralization is controlled using the method of thermal oxidative spectroscopy—oxythermography. Metals are determined by inductively coupled plasma atomic emission spectrometry. The accuracy of the analysis is controlled using the Standard Reference Material M2 moss reference sample.